Details: pc based high pc based high performance liquid chromatograph (hplc)system specification as follows: 2performance (a) solvent manager:liquid 1. pump: quaternary, gradient with flow rate range 0.01 to 10 ml/minor betterchromatogra 2. number of solvents: up to fourph 3. auto injector: 120 vial capacity auto injector (injection volume 0.1-100µl) (hplc) system 4. column oven: with two ports, 5. column temp: ambient minus 15°c or 4°c (whichever is greatest) upto 65°c6. built-in automation functions like auto start-up & shut-down, baselinestability and auto validation7. built in vacuum degassing system with thermo setting temperature up to50o c8. flow rate precision of ±0.1% rsd with flow accuracy of ±5%.9. it should include an automated software assisted purge function for easeof solvent changing and system purging/priming.10. operating pressures range up to 6000 psi for entire flow-rate range.11. needle wash system that uses fresh solvent with bef