Details: technical specifications of pacemakers, other devices and consumables for intervention (cardiology department).separate rates should be quoted for items of boq sr. no. 2 to 197 for bis/dcgi only. vvi* pacemaker with lead &accessories. all single chamber modes and basic multi-programmable parameters with preferably autosensing and autocapture/output management facilities.* must have ventricular capture management.* monitor the integrity of lead and switch polarity in case of issue.* the size of lead should be 7f or less.* the lead must be steroid eluting and should be both bipolar and unipolar configuration.* must have both active and passive fixation endocardial leads available.* company must provide at least one programmer exclusively to the department of cardiology igmc shimla.* company must provide its trained technical person for each implantation and for follow up programming whenever required.* company must quote only the latest model of devices commercially available. vvir• pace