Details: "high performance liquid chromatography (hplc): the hplc system shall include the following individual stackable self-contained modules. 1. analytical solvent delivery system: (a) it must be a high speed double micro-plunger in-parallel pump with automatic pulsation correction mechanism achieving pulse-free solvent delivery. (b) it should have a flow rate resolution of 3 nl/min(c) the flow rate should be settable between 0.0001 to 10 ml/min from micro to semi-preparative flow rates without any hardware changes (d) flow rate accuracy should be ±1% or ±0.5 µl/min of set value whichever is larger (e) flow rate precision must be less than ±0.06% rsd (f) pressure setting range should be 1-40 mpa (g) pressure pulsation must not exceed 0.3 bar (0.03mpa)(h) the gradient formation should be produced through quaternary low pressure gradient mixing. (i) the precision of composition must be less than 0.1% rsd. (j) it should employ active check valves that allow stable delivery of even non-polar or