Details: flame photometer photodiode sodium, potassium flame photometerdisplay at least 3 lines lcdflame photometer is capable to analyse elements insample:sodium,potassium,calcium,lithiumfilters: each unit of flame photometer to be suppliedwith filters for sodium and potassium, in addition theoffered unit to be supplied with filters for: sodium,potassium accuracy (+/-, in %): 0.5flame photometer shall be supplied with built in oil freeair compressor with built-in air filter with pressureregulator & gauge to deliver stable and moisture / oilfree air supply: yeswhether the flame photometer unit is offered withdigital interface along with software required for sharingsample data in digital form with computer: noaspiration rate (ml/min) 3sensitivity (sodium/ potassium/lithium : 1ppm and forcalcium: 15 ppm):yes012. flame photometer elementdirect scalescale (medical)sodium, 23 0.2 meq/1; 0.200 meq/1 ; dilution 2:40potassium,39; 0-0.1 meq/1; 0-10 meq/1; dilution 2:40lithium (optional),7; 0-0.2 meq/1;
Sector: University