Details: purchase of instruments ( laqshay) • radiant warmer- technical specification of radiant warmer with baby bassinet • a radiant warmer is used to keep the patient’s core temperature stable at 37 o c. • it should be microprocessor controlled radiant warmer with manual and servo options.• it should have facility to display both skin and air (ambient) temperature separately.. • it should have audiovisual alarm facility for overheating beyond set temperature range.• it should have alarm facility for patient temperature less than or greater than the required temperature i.e. above or below the set range.• it should rotate and swivel in different direction, so as to allow taking x-ray.• the light should be dazzle free. • it should have alarm for power failure. • it should have alarm for heater failure and auto cut off for high temp. • it should have alarm for probe failure. • it should have time out alarm in manual mode. • it should have facility to auto reset the system i