The SGPGI has published a tender for "BOQ" on the 08 Oct 2021. This tender belongs to Medical incubator category. This tender is published in Lucknow, Uttar Pradesh location. The vendors interested in this tender and related Medical incubator tenders can obtain further details by exploring Tendersniper web portal. Tendersniper sends regular tender alerts by email specifically addressing the user requirements (i.e., keywords, location and value range). Government business is a growing area of opportunity. The businesses are encouraged to actively monitor tender opportunities and participate in them to grow their business.
Related alert: incubator Tenders | medical incubators Tenders | medical equipments Tenders
Related categories: Oxygen concentrators Tenders | Oxygen concentrators maintenance Tenders | Oxygen air blender Tenders |
SGPGI issued 0 tenders for Medical incubator last year
With, 0 floated just last month.
Tender Title | BOQ |
---|---|
Tender Description | the incubator chamber volume should be 165-200 liters or more with interior components constructed of electro polished stainless steel or 100% solid copper, with rounded corners, to minimize potential for unwanted contamination and simplify cleaning.2- 2180 degree high temperature sterilization cycle with proven efficiency in contamination control against bacteria, molds, resistant bacterial spores and mycoplasma by microbiological testing and temperature mapping ensuring full chamber sterilization.3- the unit should incorporates a direct air jacketed heating design featuringhigh quality thermal jacket insulation and fan assisted circulation, allowingrecovery under 10 minutes of all parameters (temperature, co2, andrelative humidity) following a 30 second door opening. 4- the incubator should incorporate an integrated humidification design, with water reservoir that is in direct contact with a heated surface to maximize humidification efficiency evaporation.5- water level sensor in wat |
Comments | the incubator chamber volume should be 165-200 liters or more with interior components constructed of electro polished stainless steel or 100% solid copper, with rounded corners, to minimize potential for unwanted contamination and simplify cleaning.2- 2180 degree high temperature sterilization cycle with proven efficiency in contamination control against bacteria, molds, resistant bacterial spores and mycoplasma by microbiological testing and temperature mapping ensuring full chamber sterilization.3- the unit should incorporates a direct air jacketed heating design featuringhigh quality thermal jacket insulation and fan assisted circulation, allowingrecovery under 10 minutes of all parameters (temperature, co2, andrelative humidity) following a 30 second door opening. 4- the incubator should incorporate an integrated humidification design, with water reservoir that is in direct contact with a heated surface to maximize humidification efficiency evaporation.5- water level sensor in wat |
Published Date | |
---|---|
Due Date | 09 Oct 2021 00:00:00 |
Estimated Value | 0.0 |
---|---|
EMD | 0 INR |
Processing Fee | 0 INR |