Details: technical specifaction system should be based on pressure swing adsorption / vacuum swing adsorption oxygen generating technology. (make/ model should be mentioned) system should be able to produce minimum 27 m3 per hour. and system should be able to produce gas equivalent to minimum 90 or more cylinders per day of 7.3 m3 the connected electricity load should not be more than 40 kw. oxygen purity should be 93% +/- 3 (check meter/ sensor/ display should be inbuilt). supply must be medical grade oxygen only. (make/ model should be mentioned). outlet pressure should not less than 4.5kgf cm2 vertical at all times. the system should have compressor /oil free compressor (make/ model should be mentioned) discharge pressure should be more than 7 bar. air receiver & oxygen receiver storage capacity should not be less than 1000 its. in-line oxygen analyzer and line pressure measurement should be provided (make/ model should be mentioned). gas chromatography to monitor other gases should be
Sector: Medical and Health