Details: thermo gravimetric analyser(tga) specification:-i. maximum working temperature: 1200 ˚c or more ii. furace warranty: 36 months or more iii. heating rate: 0.1% ˚c/sec to 250 ˚c/sec or more iv. temperature accuracy:± 0.15 or better v. cooling time: 35 min or lesser vi. furnace type and movement: software controlled vii. crucible (alumina, copper, high pressure crucible, ceramic) must be included in the package cost. viii. calibration type: auto/manual ix. balance load capacity:1.3 g or higher x. sensitivity/resolution: 0.2 µg xi. weighing precision: ± 0.02% xii. thermostatic capability: capable xiii. software packages for running the machine and tga analyses which is compatible with windows 10 or higher version must be included in package cost. xiv. desktop computer system: branded (hp/del/equivalent brand) compatible with machine (i7 p