Details: real time pcr system • the system should be automated integrated 96 well peltier based, open system and can be upgradable to 384 well system. system should be suitable for both real-time pcr and end-point pcr analysis with sixindependently controlled zones.it should have touch screen lcd for standalone operation• onboard memory for storage of at least 1500-2000 runs with 10 gb storage capacity. . both 0.1 and 0.2 ml tube can be used and can be run by individually, by 8 tubes strips and/or 96 well plate.the minimum reaction volume should be of 10-30 μl/10-100 μl for 0.1/0.2 ml block. • the system should support the temperature range from 4°c to 100°c with block ramp more than 6°c/secthermal block should be 10.1°c temperature uniformity and 0.2°c temperature accuracy. • typical run time should be less than 40-60 minutes for 45 cycles.• system should have six de-coupled excitation and emission filter sets to enable collection of up to 21 unique combinations of wavelengths during a single
Sector: Medical Education