Details: universal testing machine for polymerspecificationmachine capacitybasic framedescription10kn or moreaccurate measuring range 200n to 10000n is mandatory.accuracy ±0.5% or betterit should be bench top type with 2 columns, 2 stiffeningrod & 2 lead screws.connections for compressed air should be built into themachine frame allowing operation of pneumatic gripswithout having long air supply lines obstructing the testarea.driveaccuracies/loadmeasurementaccuracyresolution of forcemeasurementcrossheadrangecrossheadresolutionframe should have emergency stop switch and control formachine start, stop and crosshead up-down movement.frame should have easily adjustable upper and lower travellimiting switch to set travel limit of moving crosshead inorder to protest load cells, grips, operator and machine fromany accident.the crosshead must be driven using precision ball screwscoupled with a single servo-motor. any machine with twoservo motor driving to individual lead will not beconsidered.micropr