Details: massqualitative and quantitative analysis withspectrophotometercalibration, auto tuning facilitysystemthe interface between uhplc and )interfacespectrophotometer should be capable of handlinglarge batches of samples with complex matricesover a long period of time.the cleaning and maintenance of ion source andb)de salvation line should be simple withoutbreaking vacuum.the ms should employ an angle spray which isc)ion sourceorthogonal –flow sprayer which enhances thesensitivity across a wide range of lc conditionsreduces maintenance costs.the system must be offered with dedicated esid)and apcide solvation temperature must be with stand highe)temperature up to 400 o c or morequadrupole based mass analyzer must consists off)mass analyzer2, high standard fast scanning quadrupole andone high efficiency multi pole collision cell.mass range should be from m/z 5 to 2000 amu.g)mass range0.1 da over 24 hrs or betterh)mass stability15, 000 amu/ sec or morei)scanning speed25 milli sec or lessj)pol