Details: lab pm 10 = respirable dust sampler (rds) brushless andnoiselesss.no. specification1 equipment should confirm to bis is: 5182-part-23-20062 rspm should be collected on filter paper and coarse dust should becollected in a cup under cyclone. this separation technology shouldbe based on approved design from a csir / neeri technology3 the equipment must be in use with cpcb & spcbs4 housing sturdy aluminium cabinet to house blower,filter holder assembly, time totalizer,programmable timer, flow controller & flowmeasurement device.environment protection training & research institute (eptri)91/4, gachibowli, hyderabad – 500 032, telangana – indiaphone : +91-40-23180100; 23180116,23180126email:dg@eptri.com,procurementeptri@gmail.com ; website : www.eptri.compage 19 of 295 heavy duty blower brushless and noiseless for at least 28 hourscontinuous operation.6 flow rate 0.8 to 1.4m3/min. free flow with flowstabilization by electronic feedback flow controldevice.7 filter holder aluminium casting wit