Details: supply, installation, testing and commissioning of oxygen manifold supply system (without cylinders) all copper pipe should be suitable to withstand working pressure of 145 kg/cm2. all sizes of components shall be compatable to 2x10 and 2x5 duplex manifold and existing medical gas pipeline sytem. total work shall executed as per direction of engineering - in-charge. during the execution of contract work. any item which is technically required not mentioned in boq, need to be provided by the contractor. no extra charges will be paid for such items. supply and installation of high pressure copper pipes of size 1/2" inch normal bore and 15 standard wire gauge. including clamping, pipe fittings and brazing work supply and fixing of brass through hole manifold block with lock nut to ms frame. the distance between block to block min 300 mm. compatable to 2x10 and 2x5 duplex manifolds supply and fixing of brass one side hole (end) manifold block with lock nut to ms frame. the distance betw
Sector: Medical Education