Details: state plan 2023-24 procurement of 1 no ftir at sri od94314/2024-2025:spectral range 7800–350 cm-1 mid-infrared kbr beam splitterwave length precision 0.01 cm-1 at 3,000 cm-1 or betterwave length accuracy 0.1cm-1 or betterspectral resolution: 0.6cm-1 or bettersignal to noise ratio 32,000: 1 for 1 minute or better(peak to peak) optical source mid-infrared, user replaceablelaser solid-state laserinterferometer permanently/ dynamic aligned interferometerdetector dtgs or dlatgs detectoroptics optics with sealed and desiccated optical bench with protective kbr windows. all optics must be sealed in a chassis for enhanced durability, robustness, and moisture control.diagnostics automatic accessory recognition, humidity, temperatureperformance verification automated performance verification as per certified standard ( iso 17034 compliant) and certificate should be providedsoftware suitable software for complete collection of spectral processing and analysis verification, customizable t
Sector: Water supply and Sewerage