Details: ICPMS Specifications:Liquid chromatography (LC) unit for speciation:Fully automated LC‐ICP‐MS operation for speciation studies for Methyl Mercury and Arsenic.The LC unit should consist of suitable gradient pump with integrated degasser, column compartment, auto sampler(≥100 vials) and column oven. Standards and columns for methyl mercury and arsenic to be provided. Thechromatography software module has to be integrated with ICP-MS software for real-time analysis. Statusinformation to be displayed for all vital analytical parameters of both ICP‐MS and HPLC. Complete organic solventintroduction kit, including Oxygen Mass flow Controller, Oxygen cylinder, gas panel, all related accessories &consumables including Pt cone set. IQ and OQ to be included. Five days of application training after standardinstallation and training. Nanoparticle Module along with software and standards.Additional Consumables to be included: Vial and cap 1000 No., PEEK tubing 5 m (05 sets), PFT frit (10 No.), glass