Details: bipap machine (v2)category specification bid requirement (allowed values)general product description bipap machineproduct information airflow generator pressure(ipap pressure) 4 to 25 cm of h2o or moreairflow generator pressure(epap pressure) 4 to 20 cm of h2o or moremode of operation automaticmodes available spontaneous, timed, volume assured pressure support, pac (pressure-limited assist control ventilation), spontaneous with time, manual cpaprange of back up respiratory rate(in bpm) 5 to 30 or moresensitivity settings appropriate triggering and cycling setting available for better patient synchronizationrange of minimum respiration time(sec) 0.1 to 0.3maximum respiration time(sec) 0.3 to 3.00range of rise time(in ms) 150 to 600 or morehumidifier provided yesinspiration trigger for auto start yesprovision of auto stop option when put-off mask yesprovision of inspiratory and expiratory trigger sensitivity adjustment yesprovision of inspiratory and expiratory adjustment provided yesa
Sector: Labour and Employment