Details: proposed applications: the lc-ms/ms is intended to use for the qualitative and quantitative analysis of organicmolecules like polyfluorinated alkyl substances, pesticides, phthalates, and their metabolites and other contaminantsin biological matrices.s.no. description1. a. specifications for ultra high-performance liquid chromatography system (uhplc)2.the system should have quaternary gradient pump with 1 to 4 solvents integrated with inbuilt highefficiency degassing units, minimum 4 lines with facility for auto-sampler rinsing and improved gasflow stability.3. the lc pump should have following features:4. pump quaternary gradient pump5. flow rate0.01ml to 2 ml/min or better in 0.001 ml gradient with accuracy±1% and precision 0.1% or better at 1 ml/min6. pressure range at least 15000 psi 1ml/ min or better should be offered7. purging purging of pumps automated as well as manually.8. flow accuracy ± 1%9. type of gradient linear and non-linear10. ph range 2-12 ph11. reservoir systemmobil
Sector: Medical and Health