Details: icb 17: technical specifications of gas chromatograph – high resolution mass spectrometer(gc-hrms)s. no. specification requirement1.0 instrument composition1.1 gas chromatograph one set1.2high resolutionmagnetic sector massspectrometerone set1.3 vacuum system forhrms one set1.4 cooling system forhrms one set1.5 compressed airsystem for hrms one set1.6automatic dataacquisition andprocessing unitone set1.7computer withaccessories andsoftwareone set for system control and data acquisition, another setfor post run data processing2.0 gas chromatograph2.0.1 gc systemcomputer controlled data workstation based gaschromatograph (gc). built in diagnostics and comprehensiveself testing.2.0.2 displayfunctional keyboard / touch screen display. display shallinclude temperature and pressure / flow parameters, type ofcarrier gas, carrier gas column pressure, flow rates, split flow,including method and auto sampler information.2.0.3 memory protection power fail memory protection2.0.4 storage facility u