Details: Sl No Item Qty EMD1 Centralized monitoring system for emergency ICURequired quantity: 1Purpose: To view and extract data from all 16 existing ICU multi-parameter patientmonitorsSpecifications:(i) should have facility to display upto 20 real time waves at a time(ii) should be able to be networked with atleast 16 beds.(iii) should have separate patient window for viewing detailed real-time orstored data for individual patient(iv) should have 24 hr stored patient data monitoring, 24 hr event reviewfacility, multi lead arrhythmia and ST review facility, 50 alarms strips storage,wave review(v) should support HL7 output(vi) should have option for 12L ECG monitoring(vii) should have optional facility for dual display for detailed analysis ofindividual bed without compromising on full ICU monitoring.(viii) should have facility for interfacing a laser printer for printing patientinformation and trend formats. (printer must be provided along with centralstation)(ix) should be able to visualize
Sector: Medical Education