Details: Item Name: Equipment for nasal trans membrane potential differenceUses: It will be used for measuring nasal trans membrane potential difference in children for diagnosis of cystic fibrosis .Specifications:1 It should be able measure nasal trans membrane potential difference from neonates to adults.2 It should be provided all accessories required to perform the test, measure the results and interpret the results.3 All components of the system, including the followings, should be supplied:Sl. No. Components1) Syringe Infusion pump (5 Nos.):Flow accuracy: ± 1 %LCD displayDispensing volume: 0.01 ml to 60 mlMax flow range: 500 ml/hr or moreAuto shutoff modeBattery back up at least of 2 hrsOption of bolus should be thereAlarms: obstruction, power failure, empty syringe etc.2) Powerlab (1 No.):Number of inputs: 4 or more, single ended and differentialNumber of outputs: 2 or moreMax output current: ± 50 mAAmplification: ±2 mV to ±10 V, in 12 stepsOutput ranges: ±200 mV to ±10 V, in 6 stepsMax