Details: calibration of conducted immunity testequipment 1.1: signal generator1.1.1 frequency accuracyref. frequency9 khz-500 khz (9 khz, 100khz, 500 khz)1 mhz- 2.1 ghz (1 mhz,10 mhz, 100 mhz, 500mhz, 1 ghz, 2.1 ghz)10 mhz< 0.025%< 0.2 ppm< 0.2 ppm1.1.2 rf flatness -100 dbm to 13 dbm(9 khz, 100 khz, 10 mhz,500 mhz, 1 ghz, 2.1 ghz)< 1 db1.1.3 rf output linearity -100 dbm to 13 dbm(9 khz, 100 khz, 10 mhz,500 mhz, 1 ghz, 2.1 ghz)< 1 db1.1.4 amplitudemodulation(carrier level: -20dbm, mod rate: 1khz-10 khz, 10%-90% modulationdepth)1 mhz- 2.0 ghz (1 mhz,10 mhz, 500 mhz, 1 ghz,2.0 ghz)< 0.5%1.1.5 frequencymodulation (carrierlevel: -20 dbm,mod rate: 1 khz-10khz, 10%-90%modulation depth)1 mhz- 2.0 ghz (1 mhz,10 mhz, 500 mhz, 1 ghz,2.0 ghz)< 1.5%1.1.6 2nd and 3rdharmonicsaccuracy(0 dbm, 10 dbmtest levels)9 khz-2.1 ghz (9 khz, 100khz, 500 khz,1 mhz, 10 mhz, 100 mhz,500 mhz, 1 ghz, 2.1 ghz)≤ -30 db(accuracy)1.2: spectrum analyzer1.2.1 frequency accuracy 9 khz-1 mhz (9 khz, 100khz, 200 khz,