Details: the lc pump should have following features:pump: binary/quaternary gradient pumpflow rate: 0.01ml to 1 ml/min or better in 0.001 ml gradient withaccuracy ±1% and precision 0.1% or better at 1 ml/minpressure range: at least 15000 psi 1ml/ min or better should be offeredsystem dwell volume: should be less than 400µl for total systempurging: purging of pumps automated as well as manually.flow accuracy: ± 0.5%type of gradient: linear and non-linearph range: 2-12 phreservoir system: mobile phase reservoir system to accommodate at least 2 or4 bottles each of 1 l with cork and frit.auto sampler (automated operation controllable through ms/ms software)the auto-sampler should have capacity to hold microtiter plates or multiple sample vial racksinjection loop volume range 0.1 to 50 μlcapacity of sample tray 96 vials (1.5 to 2 ml vial holder) or moretemperature range 5˚c to 40˚csample carryover <0.002% or better for caffeine or other suitable moleculecolumn oven: it must be forced air circulation
Sector: Medical and Health