Details: uv-visible spectrophotometercomplete control through pc/laptop and stand-alone operation for glp/ gmp compliance to be supplied with instrument. • true double beam optics with reference and sample side cuvette positions and aberration corrected concave blazed holographic grating in czerny – turner mounting for high energy throughput and high quality monochromatic light • the system should be of easy-to-use user interface design and should have the ergonomic touch screen display. • wide wavelength ranges of 1100 nm to 190 nm • system should be able to be upgraded to sipper pump, peltier system • high resolution 1 nm spectral bandwidth over entire wavelength range • wavelength setting and display in steps of 0.1nm • wavelength accuracy of ± 0.1nm for d2 peak 656.1 nm, ± 0.3 nm for entire range spectral line • wavelength reproducibility of + 0.1nm • wavelength slew rate 14500 nm/min or better • variable wavelength scanning speed of 29,000 nm/min or better